| General |
|
| Over Current Protection |
2X10A @ 115V and 2X5A at 230V ( Both Fuses included with Product as Accesories) |
| VA / Watt Capacity |
1000 |
| Transformer Type |
Medical Grade, Low Impedence Toroid- Line Isolation (Pass Through or Step UP or Step Down) |
| Output |
|
| Output (VA) |
1000 |
| Output (Watts) |
1000 |
| Output Receptacles |
(8) C13 |
| Output Nominal Voltage |
115V/230V (Selected by Output Switch) Can also be step up or step down, Selected when Installing Fuses. |
| Output Amp Capacity |
8.7@ 115V/ 4.3@230V |
| Input |
|
| Nominal Input Voltage Description |
115/230 VAC (Selected by Input Power Module Fuse Installation) Operating Input Voltage Range 100-125Vac or 200-250VAC |
| Input Frequency |
50/60HZ |
| Maximum Surge Amps |
70A |
| Input Current |
10A @ 115V / 5A @ 230V |
| THDi |
Less than 1% |
| Power Input Receptacle |
IEC-320-C14 |
| Surge/Noise Suppression |
|
| AC Suppression Joule Rating |
240 |
| IEEE587 Cat. A Ringwave Let-Through |
Reduced to 0.5v Common Mode |
| Noise Suppression Rating |
35-65db |
| Isolation |
|
| Output Isolation Transformer |
Toroid with AC output Floating with Respect to Earth Ground POAG Connector and Earth Ground Stud |
| Leakage Current |
Max AC Leakage Current 80 (uA) |
| Physical |
|
| Shipping Dimensions (hwd / cm) |
26.01 x 38.00 x 43.51 |
| Shipping Weight (kg) |
14.66 |
| Unit Dimensions (hwd / cm) |
13,03 x 24,99 x 30,00 |
| Unit Weight (kg) |
13.72 |
| Cooling Method |
Convection |
| Material of Construction |
Sheet Metal Cabinet |
| Color |
White |